CentriVap®micro IR Vacuum Centrifugal Concentrators Models 77010 Series Please read the User’s Manual before operating equipment.
Chapter 2: Prerequisites Product Service 1-800-522-7658 6 Location and Exhaust Requirements The CentriVap micro IR Concentrator should be located on
Product Service 1-800-522-7658 7 CChhaapptteerr 33:: GGeettttiinngg SSttaarrtteedd Now that the site for your CentriVap micro IR is properl
Chapter 3: Getting Started Product Service 1-800-522-7658 8 CentriVap micro IR Components As previously mentioned, the CentriVap micro IR System is a
Chapter 3: Getting Started Product Service 1-800-522-7658 9CAUTION: Never attempt to defeat the latch or open the lid while the CentriVap micro IR is
Chapter 3: Getting Started Product Service 1-800-522-7658 10 Chemical Resistance of CentriVap micro IR Components Your CentriVap micro IR Vacuum Conc
Chapter 3: Getting Started Product Service 1-800-522-7658 11Solvent Safety Precautions CAUTION: The CentriVap micro IR is not classified as “explosio
Product Service 1-800-522-7658 12 CChhaapptteerr 44:: UUssiinngg YYoouurr CCeennttrriiVVaapp mmiiccrroo IIRR After your CentriVap micro I
Chapter 4: Using Your CentriVap micro IR Product Service 1-800-522-7658 13 Temperature Setting Guidelines The evaporation rate achieved by the Centri
Chapter 4: Using Your CentriVap micro IR Product Service 1-800-522-7658 14 65.0 1. Display – Shows set point parameters and actual measured condit
Chapter 4: Using Your CentriVap micro IR Product Service 1-800-522-7658 15The time set point range is 00:00 (on) to 99:59 (99 hrs 59 min.) Preheat: T
Copyright © 2011 Labconco Corporation. All rights reserved. The information contained in this manual and the accompanying products are copyrighted
Chapter 4: Using Your CentriVap micro IR Product Service 1-800-522-7658 16 9. Select “OFF” or “STANDBY” to terminate operation if the CentriVap micr
Product Service 1-800-522-7658 17 CChhaapptteerr 55:: MMaaiinnttaaiinniinngg YYoouurr CCeennttrriiVVaapp mmiiccrroo IIRR Under normal op
Chapter 5: Maintaining Your CentriVap micro IR Product Service 1-800-522-7658 18 7. Check the continuity of the protective earth ground between the
Product Service 1-800-522-7658 19 CChhaapptteerr 66:: AAcccceessssoorriieess ffoorr YYoouurr CCeennttrriiVVaapp mmiiccrroo IIRR The fo
Product Service 1-800-522-7658 20 CChhaapptteerr 77:: TTrroouubblleesshhoooottiinngg Refer to the following if your CentriVap micro IR fail
Chapter 7: Troubleshooting Product Service 1-800-522-7658 21PROBLEM CAUSE CORRECTIVE ACTION Evaporation rate is reduced (continued) Obstruction
Product Service 1-800-522-7658 22 AAppppeennddiixx AA:: CCeennttrriiVVaapp mmiiccrroo IIRR CCoommppoonneennttss The following pages list
Product Service 1-800-522-7658 23 AAppppeennddiixx BB:: CCeennttrriiVVaapp mmiiccrroo IIRR DDiimmeennssiioonnss
Product Service 1-800-522-7658 24 AAppppeennddiixx CC:: CCeennttrriiVVaapp mmiiccrroo IIRR SSppeecciiffiiccaattiioonnss This Appendix cont
Appendix C: CentriVap micro IR Specifications Product Service 1-800-522-7658 25 Evaporation Rates Solvent Tube Size (ml) Number of Samples Sample S
TTAABBLLEE OOFF CCOONNTTEENNTTSS CHAPTER 1: INTRODUCTION 1 Safety Symbols 2 CHAPTER 2: PREREQUISITES 5 Electrical
Product Service 1-800-522-7658 26
CHAPTER 6: ACCESSORIES FOR YOUR CENTRIVAP MICRO IR 19 CHAPTER 7: TROUBLESHOOTING 20 APPENDIX A: CENTRIVAP MICRO IR COMPONENTS 22 APP
Product Service 1-800-522-7658 1 CChhaapptteerr 11:: IInnttrroodduuccttiioonn Congratulations on your purchase of a Labconco CentriVap micro
Chapter 1: Introduction Product Service 1-800-522-7658 2 Safety Symbols Your CentriVap micro IR Concentrator was designed with safety in mind, howeve
Chapter 1: Introduction Product Service 1-800-522-7658 3 Warning: Precautions for installation Connect to proper supply voltage. Verify that the su
Chapter 1: Introduction Product Service 1-800-522-7658 4 Do not use flammable sprays near the product. Doing so could cause fire. Protect the contr
Product Service 1-800-522-7658 5 CChhaapptteerr 22:: PPrreerreeqquuiissiitteess Before you install your CentriVap micro IR, you need to prep