INSTALLATION 6 8 Foot, 230 Volt Hoods with External Blower 230 Volt: Locate the blue wire, which is connected to the blower switch. Disconnect this
INSTALLATION 7 terminal on the white wire wiring harness #48610 by removing the terminal, stripping the wire and installing a wire nut. Locate the ye
INSTALLATION 8 increments, at intervals about 10 seconds apart to allow for delayed reaction of the alarm itself. 7. Readjust the fume hood airflow t
INSTALLATION DRAWINGS 9
INSTALLATION DRAWINGS 10
NORMAL OPERATIONS 11 Alarm Activation The audio and visual alarm will activate approximately six seconds after an alarm condition is detected. To te
ELECTRICAL WIRING DIAGRAMS 12
ELECTRICAL WIRING DIAGRAMS 13
ELECTRICAL WIRING DIAGRAMS 14
ELECTRICAL WIRING DIAGRAMS 15
Copyright © 2000, 2007 Labconco Corporation. All rights reserved. The information contained in this manual and the accompanying products are copyrig
ELECTRICAL WIRING DIAGRAMS 16
ELECTRICAL WIRING DIAGRAMS 17
ELECTRICAL WIRING DIAGRAMS 18
ELECTRICAL WIRING DIAGRAMS 19
ELECTRICAL WIRING DIAGRAMS 20
TROUBLESHOOTING 21 Symptom Suggested Recommendations No Lights Power supply not plugged into proper voltage; plug in power supply. Monitor re
TABLE OF CONTENTS Introduction Components Shipped 1 General Description 2 Performance 2 Component Identification
PREFACE Thank you for displaying confidence in us by selecting a Labconco Guardian Jr. Airflow Monitor. Our design engineers, assemblers and inspecto
INTRODUCTION 1 Components Shipped Carefully check the contents of this package for damage that may have occurred in transit. Do not discard the cart
INTRODUCTION 2 General Description The Guardian Jr. Airflow Monitor is designed to continuously monitor airflow through fume hoods. This permanently
INTRODUCTION 3 Component Identification 1. Air Inlet A portion of the air coming into the hood passes through the air inlet and across the flow sen
INSTALLATION 4 Disconnect your fume hood from its electrical service prior to beginning the installation of your new monitor. The installation of you
INSTALLATION 5 10. Insert the velocity probe into the open end of the vinyl tubing and position the probe in the 3/8" diameter hole drilled in St